Power supply Internal AC power; 115-230 V ± 10% 50/60Hz Rechargeable lead battery Pb 12 Volt, 2000 mAh
-
Battery power capacity 1 hour in continue recording, 180 recordings
-
Writing system Thermal printer, 8 dot/mm. Usable print height 210 mm
-
Thermosensitive paper Format 210 mm in gridded Z-fold packets: length 30 m., page 210x150 mm Rolls: length 17 m., page 210x280 mm
-
Display Backlit graphic display 320x240 pixels (115x86 mm) 5,7 inches
-
Keyboard Membrane, with functional and alphanumeric extended keyboard
-
Leads 12 standard leads
-
Signal memory 10 seconds each lead in automatic mode
-
Print channels 12
-
Print format Automatic mode: 3, 6×1, 6×2, “Full Page”(3×4+R)×1 and (3×4+3R)×1, 12×1 Manual mode: 3, 6, 12
-
Display channels 3, 6, 6+6
-
Operating modes Manual: acquisition and printing in real time Automatic: simultaneous acquisition Pre-programmed: in real time acquisition at programmable intervals Arrhythmia: detection of arrhythmia events (optional) HRV: acquisition and processing of heart rate variation (optional) PC ECG: transmission of signal in real time to Personal Computer (optional) Paper Saving: acquisition without print (optional)
-
Recording/display sensitivity Manual: 5 – 10 – 20 mm/mV ± 5% Automatic: according to number of channels printed
-
Paper transport speed 5 - 25 – 50 mm/s
-
Defibrillation protection Internal
-
Input dynamics ± 300 mV 0 Hz ± 10 mV band width
-
Input impedance > 100 MOhm on each electrode
-
Common mode rejection > 100 dB
-
Frequency response 0,05 - 150 Hz (–3db)
-
Time constant 3.3 s
-
Acquisition 12 bit 1000 samples/s/channel printing and filters 500 samples/s/channel in calculation and filters Resolution 5 MicroV/bit
-
Pacemaker recognition Recognizes pulse in accordance with current IEC standards
-
Filters Mains and muscle interference: modified digital notch 50 – 60 Hz Anti-drift: digital high-pass 0.5 Hz, linear phase, without morphological distortion
-
Communication interface Infrared or bluetooth*
-
Diagnostics programs Parameter calculation (optional) ECG analysis program (optional) Arrhythmia program (optional) HRV: heart rate variation (optional) Options Memory option, ECG measurements option
-
ECG analysis option
-
Arrhythmia option
-
HRV analysis option
-
PC archive option
-
PC ECG option
-
Dimensions 325 x 80 x 345 mm (length x height x depth)
-
Weight 5000 grams
-
Safety and Conformity to standards Class I, type CF Ref: EN 60601-1, EN 60601-2-25, EN 60601-2-51 According to: 93/42 EEC: class IIa * Features already available in version ar2100viewbtOptions Memory option, ECG measurements option, ECG analysis option, arrhythmia option, HRV analysis option, PC archive option, PC ECG optionDimensions 325 x 80 x 345 mm (length x height x depth Weight 5000 gramsSafety and Conformity to standards Class I, type CF Ref: EN 60601-1, EN 60601-2-25, EN 60601-2-51 According to: 93/42 EEC: class IIa